Gore DrySeal Flex Introducer Sheath, Catalog #DSF1433, 33cm 14FR, Sterile. The responsible firm on the label is W. L. Gore & Associates, Inc., Flagstaff, AZ.

FDA Recall Z-1962-2020 · Posted None

Recall StatusTerminated
Product CodeDYB
Date Initiated2019-11-18
Date Terminated2020-05-13
Product Quantity141 devices

Reason for Recall

The devices were mislabeled with the incorrect product size resulting in 16Fr devices being labeled as 14Fr devices.

Firm Action

The recalling firm issued Field Safety Notices dated 11/19/2019 with FSN Ref and FSCA Ref numbers of #0000000-11152019-001-R titled "Mislabelled 16Fr Introducer Sheaths Labelled as 14FR (3 lots)" issued via email and overnight mail. Return of the devices was requested.

Root Cause

Labeling Change Control

Distribution Pattern

All distribution is foreign to the following countries: Austria, Belgium, Czech Republic, Denmark, Finland, France, Germany, Greece, Ireland, Italy, Japan, Norway, Spain, Sweden, Switzerland, and the UK.