Philips Healthcare MobileDiagnostwDR system. Mobile Digital radiography X-Ray System Product Usage: MobileDiagnost wDR is a mobile digital radiography X-Ray System. It is intended for use by a qualified/trained doctor or technologist on both adult and pediatric patients for taking diagnostic radiographic exposures of the skull, spinal column, chest, abdomen, extremities, and other body parts. Applications can be performed with patient sitting, standing, or lying in the prone or supine positions. Not intended for mammography.

FDA Recall Z-1964-2013 · Posted 2013-08-21

Recall StatusTerminated
Product CodeIZL
Date Initiated2013-06-11
Date Terminated2017-01-12
Product Quantity240 systems US -- 1 system within Puerto Rico

Reason for Recall

The Instructions for Use (IFU) for the MobileDiagnost wDR fails to comply with a Federal standard. The IFU does not provide tolerance levels for several generator values as required by 1020.30(h)(3).

Firm Action

Phillips Healthcare sent an Important Notification letter dated July 10, 2013 to all affected customer. The notification informed customers that Philips will implement an update to the Instruction for Use (IFU). Philips will, without charge, remedy the defect or bring the product into compliance with each applicable Federal standard in accordance with a plan to be approved by the Secretary of Health and Human Services. For questions contact your local Philips representative or local Philips Healthcare office.

Root Cause

Radiation Control for Health and Safety Act

Distribution Pattern

US Nationwide Distribution including Puerto Rico