BARD Access Systems PowerPort ClearVUE Slim Implantable Port With Smooth Septum and Attachable 6 F Polyurethane Open-Ended Single-Lumen Venous Catheter, Product #5676300 - Product Usage: All materials are biocompatible, can be used with virtually all injectable solutions intended for medicinal use, including the power injection of contrast media.

FDA Recall Z-1964-2020 · Posted None

Recall StatusOpen, Classified
Product CodeLJT
Date Initiated2020-01-15
Product Quantity200 units

Reason for Recall

The catheter locks for the lot may contain an 8Fr catheter lock instead of a 6Fr catheter lock per device labeling.

Firm Action

The recalling firm issued letters dated 1/15/2020 via FedEx with proof of delivery notification. The devices were to be returned.

Root Cause

Employee error

Distribution Pattern

US Nationwide distribution including in the states of AZ, CA, MA, MD, NJ, NY, and TX. There was no foreign/government/military distribution.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.