Arrow International, Inc. Two-Lumen CVC Kit with Blue Flex Tip Catheter. The multiple lumen catheter permits venous access to the central circulation of patients.

FDA Recall Z-1965-2013 · Posted 2013-08-14

Recall StatusTerminated
Product CodeDQY
Date Initiated2013-07-31
Date Terminated2015-04-28
Product Quantity156

Reason for Recall

Arrow is recalling the product due to a labeling inconsistency. One portion of the lidstock states the kit contains a 3-Lumen catheter while the rest of the lidstock correctly identifies the catheter as 2-Lumen.

Firm Action

An "Urgent Medical Device Recall Notification" letter dated July 31, 2013 was sent to direct accounts requesting immediate discontinuation of use and quarantine of all affected product on hand. Negative responses were also requested and the immediate return of product to Arrow was authorized. Questions were directed to Customer Service at 800 343-2935.

Root Cause

Labeling Change Control

Distribution Pattern

AL, PA, CT, NC, VA, KS, RI, AZ, NJ, and IL.