TS Series Depth Gauge for Integra Ti6 Internal Fixation System 2.0/2.5mm Screws The FS2030 Depth Gauge is used by the physician to determine the correct screw length when implanting a 2.0 or 2.5mm TS Series screw for the Ti6 Internal Fixation System.

FDA Recall Z-1965-2016 · Posted None

Recall StatusTerminated
Product CodeHRS
Date Initiated2016-05-10
Date Terminated2016-12-16
Product Quantity71 units

Reason for Recall

As a result of an internal investigation of a non-conformance it has been determined: 1) Depth Gauges from Lot UP15H2 exceed the overall length specification requirements. 2) Depth Gauges from prior lot numbers manufactured before Integra acquired the product line reflect an incorrect number for the first mm depth marking on the gauge. The number should be 4, it is 6. 3) The space / holder

Firm Action

The firm, Integra Lifesciences, sent an "Urgent: Voluntary Medical Device Recall" letter dated 5/10/16 to affected consignees/customers via traceable courier service on 5/10/16. The letter describes the product, problem and actions to be taken. The customers were instructed to stop using and distributing product immediately; complete and return the Recall Acknowledgement and Return Form via fax to 1-609-275-9445 or email to [email protected], even if you do not have any of the product. Should you have any questions, please contact Customer Service at 1-866-854-8300.

Root Cause

Process design

Distribution Pattern

US Distribution to states of: CO, FL, IL, MI, NJ, NY, SC, TX, and VA.