Rapid2 PSU Stim Interconnecting Cable - Product Usage: indicated for the treatment of Major Depressive Disorder in adult patients who have failed to achieve satisfactory improvement from prior antidepressant medication in the current episode.

FDA Recall Z-1966-2020 · Posted None

Recall StatusTerminated
Product CodeOBP
Date Initiated2020-03-23
Date Terminated2024-09-20
Product Quantity71 units total

Reason for Recall

It was identified that there is a potential for the Stim Interconnecting high Voltage Cable located at the rear of the Rapid2 and Horizon power supply unit to fail.

Firm Action

The Magstim Company Ltd notified customers on about 3/23/2020 via URGENT FIELD SAFETY NOTICE Medical Device Recall letter. The letter instructed customers to identify all affected products at their facility, raise awareness to all those who need to be aware within their organization, notify customers if the product was further distributed, and complete and return the provided acknowledgement form, so Magstim Company can contact them to arrange replacement of the cable. Once customers have returned the acknowledgement form, Magstim Company will arrange for a new cable to be sent to you or your technical personnel. This will be accompanied by instructions on how to install the new cable. Magstim Technical Service Team can guide you through the installation via telephone or video call. This to limit transmission during the current pandemic COVID-19.

Root Cause

Nonconforming Material/Component

Distribution Pattern

Worldwide distribution - US Nationwide distribution including in the states of AL, AR, CA, CT, TN, IL, MN, MT, NC, NJ, NY, PA, SC, TX, WA, and the countries of Cyprus, China, Iran, Saudi Arabia, UK, Italy, Malaysia, Spain, Taiwan, Vietnam, Australia, Japan, Turkey, France and Israel.