Tearaway Introducer Kit, REF: KIT-010-12, KIT-010-17, KIT-900-16

FDA Recall Z-1968-2023 · Posted 2023-06-15

Recall StatusOpen, Classified
Product CodeDYB
Date Initiated2023-05-08
Product Quantity490

Reason for Recall

Sheath introducer assemblies could potentially, due to improper storage conditions, have a luer connection portion of the dilator hub that cracks or dislodges during use, which renders the dilator component unusable, and can require the use of a replacement device.

Firm Action

On 5/5/23, recall notices were sent to customers who were asked to do the following: 1) Identify and segregate the affected lot(s) that are in your possession as well as those in the possession of your end-users. 2) Complete and return the field correction reply form to [email protected] 2) Return affected product to the recalling firm. Customers with questions can email: [email protected]

Root Cause

Process design

Distribution Pattern

Worldwide - US Nationwide distribution in the states of PA, UT, FL, PA, OH, GA, MA, NV, CA, NY, NJ and the countries of UK, AU, TR, FR, CH, NO.