Terminated

Infinite 200 PRO in vitro diagnostic. Product Usage: The Tecan Infinite 200 PRO is a multifunctional microplate reader with injector option. The Infinite 200 PRO provides high performance for the vast majority of todays microplate applications and research and is robotic compatible. The Infinite 200 PRO has been designed as a general purpose laboratory instrument for professional use, supporting common 6 to 384 well microplates conforming to the ANS/SBS standards.

FDA Recall Z-1970-2015 · Posted 2015-07-02

Recalling FirmTecan US, Inc.
Recall StatusTerminated
Product CodeKHO
Date Initiated2015-05-29
Date Terminated2017-08-18
Product Quantity868 in total

Reason for Recall

A firmware bug causes incorrect values to be displayed for specific measured luminescence values.

Firm Action

Tecan sent an Urgent Field Corrective Action letter dated May 29, 2015 to their customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to complete the attached Customer Response Form. For questions contact your local Tecan Helpdesk.

Root Cause

Device Design

Distribution Pattern

US Nationwide Distribution in the states of AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MO, NE NH, NJ, NM, NY, NC, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, including DC

Official Source