Integre Pro Yellow LP561, Integre Pro Yellow LP6RY, Integre Pro Scan Green LP6G, Integre Pro Green LP5532 Spare, Optics Bench V2, L2G/LP6G
FDA Recall Z-1971-2023 · Posted 2023-06-27
Advertisement
Recalling FirmEllex Medical Pty Ltd.
Recall StatusOpen, Classified
Product CodeHQF
Date Initiated2023-03-16
Product Quantity57 systems in total (6 units shipped to United States)
Reason for Recall
When using a Laser-Indirect-Ophthalmoscope (LIO) as the delivery device, a scattered laser beam inside the laser cavity system may unintentionally exit through the Slit Lamp output port (SDS) and get emitted from the Objective Lens when the laser is fired.
Firm Action
Ellex Safety Bulletin is the customer notification letter. Ellex will, without charge, remedy the defect or bring the product into compliance with each applicable Federal standard in accordance with a plan to be approved by the Secretary of Health and Human Services. For questions contact your local Ellex-Lumibird representative or email us on [email protected].
Root Cause
Radiation Control for Health and Safety Act
Distribution Pattern
US nationwide/ Worldwide Distribution