Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Open, Classified

BD Texium Needle-Free Syringe: 3 mL, REF: MY8003; 5 mL, REF: MY8005; 10 mL, REF: MY8010; 20 mL, REF: MY8020; 30 mL, REF: MY8030; 60 mL, REF: MY8060

FDA Recall Z-1971-2025 · Posted 2025-06-12

Recalling FirmBD SWITZERLAND SARL
Recall StatusOpen, Classified
Product CodeFMF
Date Initiated2025-05-12
Product Quantity1,265,400

Reason for Recall

Sterile, single use closed system drug transfer devices, standalone or on infusion sets or bonded syringes, may leak at the interface between the closed male luer component and the needle-free connector, which may cause procedure delay, treatment interruptions, exposure to hazardous drugs, and/or underdosing of medication.

Firm Action

On 5/12/2025, removal notices were mailed or emailed to Medical Director, Risk Manager, Medical Device Safety Officer, Nurse Managers, and Distributors who were asked to do the following: 1. Immediately review your inventory for affected product. Destroy all unused affected products subject to the recall following your institution s process for destruction. 2. Share this recall notification with all users within your facility network of the product to ensure they are also aware of this recall. 3. Complete the Customer Response Form via email to [email protected] 4. In addition, distributors were asked to identify all customers that purchased any affected product and to provide a copy of the customer letter to all customers and to advise them of the removal notification. Report any complaints experienced with the use of this product to the firm's North American Regional Complaint Center: Phone: 1-844-823-5433 Mon-Fri 8:00am and 5:00pm CT, Email: [email protected]

Root Cause

Process control

Distribution Pattern

Worldwide - US Nationwide distribution including in the states of MA, CA, IL, AK, AL, MO, FL, VA, ME, NC, MD, IA, WI, AZ, CO, KY, IN, TX, AR, NJ, NY, MT, NE, TN, PA, GA, WA, KS, MN, ID, SC, MI, SD, OH, WV, LA, ND, VT, GU, NV, UT, OK, OR, DE, NM, MS, WY and the countries of CA, AU, NZ, CN.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑