Single-Lumen Central Venous Catheterization Kit with Blue FlexTip(R) Catheter and Sharps Safety Features; Product Codes: ASK-04301-TJ

FDA Recall Z-1974-2018 · Posted None

Recall StatusTerminated
Product CodeDQY
Date Initiated2018-04-11
Date Terminated2020-05-15
Product Quantity416,055 total products

Reason for Recall

Product sterility may be compromised due to unsealed packaging.

Firm Action

On April 11, 2018 Arrow International mailed Urgent Medical Device Recall notifications to affected customers. Customers were instructed to take the following actions: 1. If you have affected stock, immediately discontinue use and quarantine any products with the product codes and lot numbers listed above, so that the affected products can be returned to Arrow. 2. To return product, complete the enclosed Recall Acknowledgement Form and fax it to 1-855-419-8507, Attn: Customer Service or email to [email protected]. This will allow us to document the amount of product you have on hand for return. A customer service representative will contact you with a Return Goods Authorization (RGA) Number and will provide instructions for the return of products to Arrow International. 3. If you have no affected stock, please complete the enclosed Recall Acknowledgment Form and fax it to 1-855-419-8507, Attn: Customer Service or email to [email protected]. This will allow us to document your receipt of this letter. Customers with questions may contact your local sales representative or Customer Service at (866)246-6990.

Root Cause

Under Investigation by firm

Distribution Pattern

US Nationwide, AUSTRALIA, AUSTRIA, BELGIUM, CANADA, CZECH REPUBLIC, DENMARK, ECUADOR, FRANCE, GERMANY, GREECE, HONG KONG, INDONESIA, IRELAND, ITALY, JAPAN, KOREA, NETHERLANDS, NEW ZEALAND, PANAMA, PHILIPPINES, POLAND, PUERTO RICO, ROMANIA, SAUDI ARABIA, SINGAPORE, SLOVENIA, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, THAILAND, TRINIDAD AND TOBAGO, UNITED KINGDOM