Terminated

TAPSYSTEM Model 2A, REF 8002A

FDA Recall Z-1976-2019 · Posted None

Recalling FirmCardiocommand Inc.
Recall StatusTerminated
Product CodeLPA
Date Initiated2019-06-24
Date Terminated2021-04-01
Product Quantity6 units

Reason for Recall

There is a potential for an assembly error resulting in the battery wire being connected to the ON/OFF power switch incorrectly.

Firm Action

On June 24, the firm plans to send URGENT MEDICAL DEVICE letters to six consignees that will advise them of the issue, request that they either return their device or perform an ON/OFF switch verification procedure, and to return the attached REPLY FORM.

Root Cause

Process control

Distribution Pattern

United States - AR, NM, PA, TX, and WA.