Endurant II/IIs Stent Graft System: STENT GRAFT ETBF3216C145E ENDUR II BIF. For treatment of vascular aneurysms.
FDA Recall Z-1976-2021 · Posted None
Advertisement
Recalling FirmMedtronic Vascular, Inc.
Recall StatusTerminated
Product CodeMIH
Date Initiated2021-04-28
Date Terminated2024-04-05
Product Quantity3 devices
Reason for Recall
Incorrect stentstop assembly (18fr) was used to build the delivery system component of the device. Correct stentstop assembly is 20fr.
Firm Action
Beginning on 04/28/2021, the Recalling Firm initiated a verbal notification and retrieved affected product for a single U.S. customer. On 05/10/2021, the Recalling Firm sent an "URGENT MEDICAL DEVICE RECALL" Notification Letter via UPS 2-day delivery to inform its impacted customer that the Recalling Firm has determined that the stent graft system was manufactured with an incorrect stentstop assembly. The letter was also to to ask for confirmation that the customer no longer had the affected devices in its possession.
The letter instructed customers to take the following actions:
1. Complete the enclosed Customer Confirmation Form and email to [email protected].
2. Forward this notice to all those who need to be aware within your organization.
If there are any questions - contact the Recalling Firm's Field Representative.
Root Cause
Process control
Distribution Pattern
U.S. Nationwide distribution in the state of GA. O.U.S.: None
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.