Humeral Augmented Tray (HAT) trials that are components of KIT-321HAT. Each tray contains 1 of each of the following catalog numbers for a total of 6 trials/tray: (1) HAT STD Left Tray Trial, Cat. #321-00-01; (2) HAT STD Right Tray Trial, Cat. #321-00-02; (3) HAT LAT Left Tray Trial, Cat. #321-00-11; (4) HAT LAT Right Tray Trial, Cat. #321-00-12; (5) HAT EXT LAT Left Trial, Cat. #321-00-21; and (6) HAT EXT LAT Right Trial, Cat. #321-00-22. The Humeral Augmented Tray is designed to function with the Equinoxe Reverse Shoulder System orthopedic implant components.

FDA Recall Z-1977-2021 · Posted None

Recalling FirmExactech, Inc.
Recall StatusCompleted
Product CodePHX
Date Initiated2021-05-11
Product Quantity36 devices

Reason for Recall

The HAT trial contains a retaining ring that has shown the potential to disengage during use.

Firm Action

The firm issued letters dated 5/11/2021 explaining the issue, the clinical impact, and instructing the consignee to immediately cease distribution or use of the product. The information was to be extended to their accounts who may have the product in their possession.

Root Cause

Device Design

Distribution Pattern

US Nationwide distribution in the states of AZ, CO, FL, NY, SC, and TN. There was no foreign/government/military distribution.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.