Terminated

CHAMPION Medical Recliner, AC Powered with heat and massage, one chair per carton, or one chair wrapped in a blanket if delivered in person. Recliner provides heat and massage after surgery.

FDA Recall Z-1980-2013 · Posted 2013-08-17

Recalling FirmInvacare Corporation
Recall StatusTerminated
Product CodeINO
Date Initiated2013-06-18
Date Terminated2014-01-09
Product Quantity285 units

Reason for Recall

There is a possibility of elevated temperatures in the seat pad. In addition, the hospital cord set does not agree with the labeling (information in manual).

Firm Action

Champion Manufacturing plans to initiate this recall within approximately 10 business days after the submission of this report through a phone campaign and notification letters.

Root Cause

Process control

Distribution Pattern

Nationwide distribution: USA including states of: CA, FL, IN, MN, NJ, NY, OH, RI, SC, SD, TX, and VA.