Terminated

Elekta Unity, Image-Guided Radiation Therapy System Elekta Unity using Magnetic Resonance Imaging is indicated for radiation therapy treatments and stereotactic radiation treatments of malignant and benign diseases anywhere in the body as determined by a licensed medical practitioner in accordance with a defined treatment plan.

FDA Recall Z-1980-2019 · Posted 2019-07-16

Recalling FirmElekta, Inc.
Recall StatusTerminated
Product CodeIYE
Date Initiated2019-06-24
Date Terminated2021-04-01
Product Quantity12 units

Reason for Recall

The QA software solution to perform the MR to MV alignment check, does not display the stored MR to MV offset values. The user is unable to independently inspect the values during their QA.

Firm Action

The firm, Elekta, disseminated their "URGENT IMPORTANT FIELD SAFETY NOTIFICATION" by email on 06/24/2019 to its customers. The notices identified the problem, provided recommended actions, and requested the notice be placed in a place accessible to all users. Customers were also instructed to complete and return the Acknowledgement Form. A software update will be provided once available. If you have any questions, contact the Post Market Surveillance Specialist by email: [email protected].

Root Cause

Software design

Distribution Pattern

US Distribution to states of: TX & WI, and Internationally to: Canada, Denmark, Germany, Italy, Netherlands, Sweden, and the UK.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →