SmartBlock PainPump, Part numbers: P49524-S, P49534-S, P49624-S, and P49634-S.

FDA Recall Z-1981-2011 · Posted 2011-04-14

Recall StatusTerminated
Product CodeMEB
Date Initiated2010-03-17
Date Terminated2011-04-25
Product Quantity368 units

Reason for Recall

Elastomeric tube in product's compression unit has the potential to burst following filling.

Firm Action

Firm notified consignees by e-mail and telephone on 3/17/2010. The correspondence identified the affected product and lot numbers; and asked that customers specify which parts, lot numbers and in what quantities they still had in their possession.

Root Cause

Other

Distribution Pattern

Nationwide Distribution

Related FDA 510(k) Records