Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Terminated

Moore Hip Prosthesis Long Fenestrated Stem, several head Diameter, 203 mm Stem length. Used to seat the prosthesis with or without bone cement.

FDA Recall Z-1981-2013 · Posted 2013-08-17

Recalling FirmZimmer, Inc.
Recall StatusTerminated
Product CodeJDG
Date Initiated2013-07-16
Date Terminated2015-04-30
Product Quantity1,710 units are in distribution

Reason for Recall

Zimmer recently conducted a review of historic packaging validations completed. Based on this review, it was determined that a subset of product packaged at a single Zimmer facility required testing to ensure that previously packaged product from this facility met the established requirements for sterile barrier integrity. This testing determined that a single packaging tray configuration, which

Firm Action

Zimmer is removing the devices from the field including distributor and hospital inventories. A notification was issued to known users of the product on 07/16/2013.

Root Cause

Packaging process control

Distribution Pattern

Worldwide Distribution-USA (nationwide) including the states of CT, FL, GA, IL, IN, KS, KY, MA, MI, MN, MT, NC, NJ, NY, OH, OR, PA, SC, TN, TX, UT, VI, and WI, and the countries of Singapore, Canada, Mexico, Arab Emirates, Saudi Arabia, and Hong Kong.

Official Source

View FDA Recall →