Terminated

Product Name: nanoPOINT Injector System, item number 37-0000-08. Intended to fold and insert a STAAR Surgical Collamer nanoFLEX or Afinity IOLs, through surgical procedure in a human eye.

FDA Recall Z-1984-2013 · Posted 2013-08-20

Recalling FirmMedicel Ag
Recall StatusTerminated
Product CodeHQL
Date Initiated2013-08-05
Date Terminated2014-03-11
Product Quantity68 boxes or 680 units

Reason for Recall

The firm initiated this recall of one lot (121402) of nanoPOINT Injectors after they noticed that Tyvek lids might have been damaged and, therefore, the stability of the sterile barrier might be negatively affected.

Firm Action

Medicel AG sent recall notification to its one direct consignee via email on 08/05/2013. The recall notification letter was titled "URGENT-Medical Device Correction Notice for nanoPOINT Injectors LOT 121402". The recall notification letter informed the customer of the product device description and lot number recalled, reason for recall, instructions to "not use this affected lot...and return them", contact information and an area on the bottom of the letter to acknowledge receipt of the correction notice. The firm's direct consignee who further distributed this product to the end users will send out notifications.

Root Cause

Use error

Distribution Pattern

Nationwide Distribution.