Terminated

Siemens Mobilett XP Digital Mobile X-ray System Product Usage: The Mobilett XP and Mobilett XP Hybrid are radiographic systems designed for use in wards, intensive care, and premature birth-wards, pediatric and emergency departments, operating theatres as well as the central X-ray department.

FDA Recall Z-1985-2014 · Posted 2014-07-02

Recall StatusTerminated
Product CodeIZL
Date Initiated2014-05-09
Date Terminated2014-12-15
Product Quantity10

Reason for Recall

There is a potential fire hazard of certain Lithium-Ion batteries in the computers used in the original manufacturing or as spare parts between October 2010 and April 2011 for the Siemens Mobilett XP Digital mobile X-ray System, which might occur at any time with higher probability while the system is being charged and connected to the main power supply.

Firm Action

Siemens sent a Safety Advisory Notice letter dated May 9, 2014, to affected customers. The letter identified the affected product, problem and actions to be taken. The letter stated that a Siemens service technician would be visiting their site to identify and service affected devices.

Root Cause

Component design/selection

Distribution Pattern

US Nationwide Distributions in the states of MO, GA, FL, NJ, CA, PA and including NM.