Philips HeartStart MRx M3536A Defibrillator/Monitor with Software R.03.03 The Heartstart MRx is for use for the termination of ventricular tachycardia and ventricular fibrillation. The device is for use by qualified medical personnel trained in the operation of the device and qualified by training in basic life support, advanced cardiac support, or defibrillation. It must be used by or on the order of a physician

FDA Recall Z-1986-2012 · Posted 2012-07-11

Recall StatusTerminated
Product CodeMKJ
Date Initiated2012-06-20
Date Terminated2013-08-14
Product Quantity35 units

Reason for Recall

Defibrillator/monitor cannot analyze an ECG report during acquisition of 12 lead ECG

Firm Action

Philips Healthcare issued an Urgent Voluntary Medical Device Correction letter on June 20, 2012, to all affected customers. The Notification informs customers of the issue, identifies details (including models and software versions applicable) of units affected, instructions on actions to take while awaiting the correction, and identifies what action Philips plans to eliminate or remedy the issue A Philips Healthcare representative will contact customers to arrange for the updated R.03.03 software installation in all the affected devices. For questions customers were advised to call 1-800-722-9377. For questions regarding this recall call 978-687-1501.

Root Cause

Nonconforming Material/Component

Distribution Pattern

Worldwide Distribution - USA including FL, IN, MA, M, NY, OH, TN, TX, and VA. Internationally to AUSTRALIA, HONG KONG, and INDIA