Terminated

IntelliSpace Portal software and Extended Brillance Workspace software Intellispace Portal is a multimodality thin-client applications server that delivers full diagnostic viewing and clinical applications to the enterprise. The Philips Extended Brillance Workspace is intended for use as a diagnostic viewing and processing workstation.

FDA Recall Z-1987-2012 · Posted 2012-07-11

Recall StatusTerminated
Product CodeJAK
Date Initiated2012-05-23
Date Terminated2013-09-30
Product Quantity38

Reason for Recall

The Intellispace Portal software and Extended Brillance Workspace software may have significantly different results in the colors and size of the affected area when the customer uses the "Time Sensitive" and "Time Insensitive" methods.

Firm Action

Philips Healthcare sent a Urgent Medical Device Correction letter dated May 31, 2012, to all affected customers. The letter was sent via certified mail to each customer. The letter identified the product the problem and the action needed to be taken by the customer. Customers were instructed to review the information contained in the letter with all members of their staff who need to be aware of the contents of the communication. For any further information or support, customers can contact their local Philips representative or local Philips Healthcare office at (1-800-722-9377 Option 5 : Entersite ID or follow the prompts). Additionally, a Field Service Engineer will contact each site to schedule time to implement the Field Change Order. The user was informed that a field service representative will visit the facility to install the new software. Philips apologizes for any inconveniences caused by this problem.

Root Cause

Software design

Distribution Pattern

Worldwide Distribution -- Nationwide Distribution including the states of AZ, OR, PA and TX and the countries of Australia, Belgium, Canada, Denmark, Finland, France, Germany, India, Indonesia, Israel, Russia, South Africa, South Korea, Spain, Sweden, Switzerland, and United Kingdom.