Biomet 3i Certain-compatible 4.1 mm Milled Ti Abutment REF 9001767-F Preat milled Abutments are intended to be used in conjunction with endosseous dental implants in the maxillary or mandibular arch to provide support for single-unit or multi-unit prosthetic restorations.
FDA Recall Z-1987-2025 · Posted 2025-06-18
Advertisement
Recalling FirmPreat Corp
Recall StatusOpen, Classified
Product CodeNHA
Date Initiated2025-04-02
Product Quantity145 abutments
Reason for Recall
Due to manufacturing issues, abutment products were manufactured with screw seat located too high, which may cause the mating screw to have a shorter engagement length into the associated implant.
Firm Action
On 04/02/2025 an initial customer notification was distributed by the firm. On 05/05/2025, the firm sent via certified mail an updated "URGENT: MEDICAL DEVICE RECALL UPDATE" Letter to customers informing them that due to manufacturing issues with a dimensional condition that may lead to no engagement between the prosthetic screw and the implant.
Customer are instructed to discard or return the affected devices to PREAT Corporation.
For questions, contact Quality Assurance at 800-232-7732 or email [email protected]
Root Cause
Process control
Distribution Pattern
U.S. Nationwide distribution in the states of AZ, CA, GA, MO, NJ, NY, OH, PA, RI, TN,TX, and WA.