Excelsior Disposable Syringe Heparin Lock Flush, USP syringe (5 mL Fill in 10 mL syringe). For flushing of IV catheters and IV tubing only.

FDA Recall Z-1989-2016 · Posted 2016-06-14

Recall StatusTerminated
Product CodeNZW
Date Initiated2016-04-28
Date Terminated2017-08-08
Product Quantity184,500

Reason for Recall

Certain lots of Heparin Lock Flush, USP Syringes (5 ml Fill in 10 ml Syringe) failed Anti-Factor IIa assay for potency during testing of retention samples for heparin potency.

Firm Action

Excelsior Medical Inc. sent a recall letter on 4/29/2016 via certified mail. Customers are instructed to discontinue using the product subject to this recall. Customers are requested to complete the customer reply form and fax it (866) 807-3714, so arrangements can be made to return the product to Excelsior.

Root Cause

Nonconforming Material/Component

Distribution Pattern

US Nationwide distribution.