5008X HDF Hemodialysis Blood Tubing Set, 24 units per case, with the following models: a. 5008X Standard HDF Bloodline, Model Number: 03-5100-7. b. 5008X Standard HDF Bloodline w/CLIC, Model Number: 03-5100-7C. c. 5008X Standard HDF Bloodline w/ Twister, Model Number: 03-5150-2.
FDA Recall Z-1992-2025 · Posted 2025-06-18
Recalling FirmFresenius Medical Care Holdings, Inc.
Recall StatusOpen, Classified
Product CodeKDI
Date Initiated2025-04-29
Product QuantityModel Number, Cases (Eaches): 03-5100-7, 76 (1,824); 03-5100-7C, 5 (120); 03-5150-2, 8 (192)
Reason for Recall
Potential of blood leak occurring between the lower port of the venous chamber and the tubing include: hemodiafiltration (HDF), hemodialysis (HD), hemofiltration (HF), and isolated ultrafiltration.
Firm Action
Fresenius Medical Care notified its single consignee on 04/29/2025 via telephone call. The consignee was made aware of the issue and affected units and instructed to examine their inventory for those affected units. If they had any affected units, they were instructed to remove them from their inventory immediately and provide contact information to arrange for their return.
On 05/14/2025, Fresenius Medical Care expanded the recall to include additional lot numbers and notified consignees. The notification was via telephone call and included the same instructions as the initial letter but included the additional lots.
On 06/05/2025, Fresenius Medical Care expanded the recall again to include additional lot numbers and notified consignees. The notification was via telephone call and included the same instructions as the initial letter but included the additional lots. A recall notification letter was sent to the consignee on about 06/23/2025. The letter provided the same information as was verbally communicated in notifications.
Root Cause
Process control
Distribution Pattern
US Nationwide distribution in the state of Massachusetts.