Flexichamber Anti-Static Valved Collapsible Holding Chamber, used by patients to administer aerosolized medication from most pressurized Metered Dose Inhalers(pMDIs).
FDA Recall Z-1995-2016 · Posted 2016-06-15
Advertisement
Recalling FirmFSC Laboratories, Inc
Recall StatusTerminated
Product CodeNVP
Date Initiated2016-05-16
Date Terminated2017-02-10
Product Quantity1435 units
Reason for Recall
Product had torn/loose seal around the mouthpiece.
Firm Action
Letters were sent to consignees on 5/16/2016 informing them of the recall and requesting that they immediately discontinue use of the product. The letter included a reply form to be returned to FSC Laboratories.
Root Cause
Process control
Distribution Pattern
US Nationwide distribution.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.