EVARREST Fibrin Sealant Patch, Manufactured by: Omrix Biopharmaceuticals, Ltd, Israel. Indicated for use with manual compression as an adjunct to hemostasis for control of bleeding during adult liver surgery and soft tissue bleeding in adults when control of bleeding by standard surgical methods of hemostasis is ineffective or impractical.
FDA Recall Z-1996-2016 · Posted 2016-06-15
Advertisement
Recalling FirmEthicon, Inc.
Recall StatusTerminated
Product CodeGEL
Date Initiated2016-04-11
Date Terminated2017-08-02
Product Quantity228 units
Reason for Recall
Ethicon is recalling EVARREST Fibrin Sealant Patch because the characters "0-bad" were printed in the expiration date field on the foil pouch labels of the Fibrin Patch units instead of the valid expiration date (2016-12-28).
Firm Action
Ethicon sent an "Urgent: Product Recall (Removal)," dated April 14, 2016 via UPS next day mail to their affected customers. The letter identified the affected product and the reason for the recall. The letter also states that product from the affected lot is still within the expiration date, and health care practitioners, who have treated patients with the product should continue to follow patients in the usual manner. Users are to examine their inventory and quarantine the affected product. The affected product should be returned per the instructions outlined in the letter. Ethicon will provide replacement product upon receipt. Customers are to complete and return the Business Reply Form. If customers need assistance returning the product, they can contact Stericycle at 1-866-761-9472. If there are questions regarding the replacement product, customers can contact Ethicon Customer Support Center at 1-877-ETHICON.
Root Cause
Error in labeling
Distribution Pattern
Nationwide Distribution
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.