Terminated

STANLEY Healthcare Arial 54315 Network Manager

FDA Recall Z-1996-2019 · Posted None

Recall StatusTerminated
Product CodeIQA
Date Initiated2019-06-14
Date Terminated2021-05-13
Product Quantity137 units

Reason for Recall

STANLEY Healthcare has received reports of outages on the Arial nurse call system. These outages have been traced back to the Arial 54315 Serial-to-IP Network Gateway.

Firm Action

The firm initiated the recall by email on 06/14/2019. The letter requests that all users that receive a Warning flag on an Arial computer immediately contact the Technical Support Department to provide assistance in troubleshooting. The firm will be distributing replacement units

Root Cause

Under Investigation by firm

Distribution Pattern

Nationwide and Canada