Physician Order Form (PL-15400-00, Nov. 2018) distributed to end user used with the following: Prometra Programmable Infusion Pump System: MODEL/CATALOG #(s): Catalog No. US Model GTIN-14 11827 Prometra Programmable 20mL Pump 00810335020082 13827 Prometra II Programmable 20mLPump 00810335020228 16827 Prometra II Programmable 40mLPump 00810335020242
FDA Recall Z-1996-2021 · Posted None
Advertisement
Recalling FirmFlowonix Medical Inc
Recall StatusTerminated
Product CodeLKK
Date Initiated2021-05-26
Date Terminated2024-02-20
Reason for Recall
Physician Order Form (PL-15400-00 Nov 2018) distributed is discontinued due to the text 'Refill the pump and program refill with the same medication removed prior to MRI' may be misinterpreted as 'reuse the same drug solution extracted from the pump prior to the MRI procedure', creating a possible risk of infection
Firm Action
Flowonix Customer letters) have been mailed via USPS 2 day priority mail with electronic tracking on 5/26-5/27. Letter states reason for recall, health risk and Actions to be taken by the Customer:
1) Discontinue the use and distribution of the Physician Order Form (PL-15400-00, Nov.
2018). This form is no longer available for use and shall no longer be used.
2) Complete and return the attached response form to Flowonix.
Should you have any questions, please contact your Flowonix
Representative or our Technical Solutions Department at (855) 356-9665
Root Cause
Error in labeling
Distribution Pattern
US Nationwide distribution.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.