Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Open, Classified

iLet Bionic Pancreas, REF: BB1001

FDA Recall Z-1996-2026 · Posted 2026-04-29

Recalling FirmBeta Bionics, Inc.
Recall StatusOpen, Classified
Product CodeQFG
Date Initiated2026-03-12
Product Quantity1,080

Reason for Recall

insulin pump software versions 1.4.2, and 1.4.3 issue with Dexcom Continuous Glucose Monitoring (CGM) G7 sensor may result in delayed CGM reading then switch to non-CGM dosing, leading to only basal/requested meal doses provided, no correction doses/insulin reductions/suspensions for high/low glucose unless blood glucose entered for sensor life or replacement. Potential health risks to users include severe hyperglycemia with and without DKA and severe hypoglycemia.

Firm Action

On 3/12/2026, correction notices were emailed to customers who were asked to do the following: Update device software to host software 1.4.4 and up. Update your software 1. Download or update your iLet app on your phone. Read the iLet App User Guide: https://www.betabionics.com/wp-content/uploads/LA000056_F-iLet-Mobile-App-User-Guide.pdf 2. Log into your iLet app. 3. Position your iLet next to your phone with the app open. 4. Go to app settings on the iLet app and follow the on-screen instructions to update to the latest firmware. -Distributors are asked to provide patient contact information so the firm can provide the correction notice to these patients. - Complete the acknowledgment survey https://sprw.io/stt-6HnQU If you have any questions, please call the firm's Customer Care at 1-855-745-3800 option 1.

Root Cause

Software change control

Distribution Pattern

US Nationwide distribution in the states of TX, PA, AL, MI, OH, FL, TN, GA.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑