PVT-681MV Endocavitary Transducer used with the Aplio XG and Xario XG Ultrasound Systems.

FDA Recall Z-1997-2010 · Posted 2010-07-13

Recall StatusTerminated
Product CodeITX
Date Initiated2010-01-05
Date Terminated2012-02-29
Product Quantity27

Reason for Recall

Toshiba America Medical Systems is initiating a field correction on the PVT-681MV Endocavitary Transducer used with the Aplio XG and Xario XG Ultrasound Systems due to an incorrect sweep direction setting. Images acquired using the endo-cavitary transducer PVT-681MV and cross-sectional images created from these volume images are reversed horizontally on the Aplio XG (SSA-790A) and the Xario XG (

Firm Action

Toshiba America Medical Systems (TAMS) will issue notification to consignees via USPS return receipt mail. The letter contains a return reply form that is to be faxed to TAMS as proof of notification. matter. Questions regarding this letter are directed to the firm at (800) 421-1968 or to your local Toshiba representative.

Root Cause

Other

Distribution Pattern

Nationwide