Hemoccult ICT (Immunochemical Fecal Occult Blood Test) is a rapid, visually read, qualitative immunochemical chromatographic method for detection of human hemoglobin in fecal samples.
FDA Recall Z-1997-2013 · Posted 2013-08-22
Advertisement
Recalling FirmBeckman Coulter Inc.
Recall StatusTerminated
Product CodeKHE
Date Initiated2013-07-17
Date Terminated2014-10-24
Product Quantity68,116 units
Reason for Recall
Beckman Coulter has received a customer complaint that suggests trending of
positive results has increased from approximately 4% to 1 0%. As a result of our
investigation, Beckman Coulter has confirmed an increase in sensitivity at the
borderline positive level.
" This testing also demonstrated that the issue does not affect results reported as
negative.
Firm Action
Beckman Coulter sent an Urgent Product Correction letter dated July 9, 2013, to all affected customers. The letter was accompanied with a cover letter and response form, dated July 9, 2013.
Customers were informed of the product under recall with product description, actions to take, contact information, the issue, impact, and resolution.
Beckman Coulter conducted a risk assessment for this issue. Per the risk documentation, there is only a remote probability that use of the product will cause any adverse health consequences,and the product is not likely to lead to serious adverse health consequences.
If you have any questions please call (714) 961-3777.
Root Cause
Under Investigation by firm
Distribution Pattern
Worldwide Distribution - USA and the country of Canada.