Devon Supine, Lithotomy and Trendelenburg Positioning Kit; Product Code: 31151090 Product Usage: The device is intended to provide padding under the patient, and help maintain the position of an anesthetized patient during surgery.
FDA Recall Z-1997-2014 · Posted 2014-07-07
Advertisement
Recalling FirmCovidien LLC
Recall StatusTerminated
Product CodeCCX
Date Initiated2014-05-08
Date Terminated2015-09-04
Product Quantity264 devices
Reason for Recall
Straps may separate from the foam pad.
Firm Action
Covidien Sales Representatives notified customers in person by beginning May 8, 2014. The Sales Representative will provided a copy of the Recall letter (attachment 4), and collect any remaining product at the site. The notification and recovery of the product is expected to be complete no later than May 14, 2014.
Root Cause
Process control
Distribution Pattern
US Nationwide Distribution in the states of NJ, CA, TX, IL, MO, PA, OH, IN, MI, MN, FL, IN, DC, PA, VA, MS, ME, NV, AR, KY, NY, KS, SC
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.