300 Microliter CO-RE Tips for use with Microlab STAR, manufactured by Hamilton Company, Reno, NV. Product is a disposable product used with Microlab STAR series Instruments, which are microtiter diluting and dispensing devices.
FDA Recall Z-1999-2009 · Posted 2009-09-09
Advertisement
Recalling FirmHamilton Co
Recall StatusTerminated
Product CodeJTC
Date Initiated2009-04-09
Date Terminated2010-01-04
Product Quantity20 cases of Part 235950 and 27 cases of Part 235902.
Reason for Recall
Product Defect-- some tips from the mold cavity 28 are too short and do not always aspirate or dispense liquid correctly.
Firm Action
The firm initiated the recall on April 10, 2009, and issued notification to consignees via certified mail and e-mail. Customers were instructed to return the affected product to the Hamilton Company. A fax-back form was supplied with each notification. Direct questions about the recall to the Hamilton Company by calling 1-775-858-3000, extension 236.
Root Cause
Process control
Distribution Pattern
Worldwide Distribution -- US (states of DE, NJ, PA, MA and CA), New Zealand and Japan.