Roche ACCU-CHEK Performa blood glucose test strips, Roche Diagnostics, Indianapolis, IN. Test strips for quantitative blood glucose measurements in fresh capillary, arterial, or neonatal blood, as well as in heparin (lithium or sodium) or EDTA anticoagulated venous blood.

FDA Recall Z-2005-2009 · Posted 2009-09-16

Recall StatusTerminated
Product CodeNBW
Date Initiated2009-06-10
Date Terminated2010-10-29
Product Quantity1,518 of total products

Reason for Recall

There may be a significant negative bias when testing neonate samples.

Firm Action

Roche Diagnostics notified consignees in the affected countries of the problem by "Urgent Field Safety Notice" dated June 10, 2009 and instructed to cease using the Accu-Chek Performa and Accu-Chek Inform II systems on neonates, except for systems using test strips with advanced chemistry. For questions and additional information, contact your local Roche Diagnostics sales affiliate or sales representative.

Root Cause

Process control

Distribution Pattern

International Distribution -- Argentina, Australia, Brazil, Chili, China, Colombia, Ecuador, France, Germany, Great Britain, Hong Kong, India, Korea, Malaysia, Mexico, New Zealand, Pakistan, Panama, Peru, Philippines, Singapore, South Africa, Taiwan, Thailand, Uruguay and Venezuela.