PushTracker E3 component utilized with the SmartDrive MX2+ Power Assist Device (Part Numbers: MX2-32K, MX2-32P, MX2-32PK, MX2-33D, MX2-33P, MX2-33PK, MX2-33S, MX2-150, MX2-167, MX2-32D); a wearable watch that communicates with a SmartDrive Power Assist Device via the SmartDrive MX2+ Application

FDA Recall Z-2005-2023 · Posted 2023-06-26

Recalling FirmPermobil
Recall StatusOpen, Classified
Product CodeITI
Date Initiated2023-05-10
Product Quantity6,196 units

Reason for Recall

When multiple processes are running on the watch's Central Processing Unit (CPU), the application may fail unexpectedly. If this happens, the motor on the power assist device continues to run and the user may not be able to stop the device using tap gestures.

Firm Action

Permobil began notifying customers via in-person and through letters on 05/10/2023. Customers were instructed to examine their device and update their software to SmartDrive MX2+ Version 1.1.00, to acknowledge the update was conducted and the notification letter was understood through the field action portal, notify customers if the affected devices were further distributed. They were also instructed to discontinue using the affected device while waiting for the correction.

Root Cause

Software design

Distribution Pattern

Worldwide distribution.