Terminated

HACH SteriChek Residual Peroxide Reagent Strips, 5/100-strip bottles per kit, Kit Part #811905, Bottle part #1905LB, Hach Co., Ames, IA. SteriChek Residual Peroxide Reagent Strips provide a convenient means of measuring the concentration of peroxide remaining in water being used to rinse out dialysate lines and dialyzers following disinfection with peroxide/peracetic acid.

FDA Recall Z-2016-2009 · Posted 2009-09-04

Recalling FirmHach Co
Recall StatusTerminated
Product CodeLIF
Date Initiated2009-04-21
Date Terminated2009-10-07
Product Quantity275 kits

Reason for Recall

Some of the reagent strip bottles contain the incorrect strips

Firm Action

Hach Company issued an "Urgent Medical Device Recall" letter dated April 21, 2009 explaining the reason for recall and requested product remaining in inventory be returned. Subrecall to the end-user level was requested. A Return Form was also enclosed to be completed to indicate the amount of product being returned. Between April 22-23, 2009, the firm telephoned their customers to assure they received the recall letter. For further information, contact Hach Customer Service at 1-800-548-4381 extension 3560.

Root Cause

Labeling mix-ups

Distribution Pattern

Nationwide Distribution -- NJ, CA, IN, PA, NV, CA, and ME.