AXIOM Artis BA, Model Number 5904656
FDA Recall Z-2016-2019 · Posted None
Advertisement
Recalling FirmSiemens Medical Solutions USA, Inc
Recall StatusTerminated
Product CodeOWB
Date Initiated2019-06-12
Date Terminated2020-09-18
Product Quantity7
Reason for Recall
A tolerance issue in the power supply of the generator control above a specific value may cause generator A100 failure. As a result, the X-ray tube will not be supplied with the required voltage and X-ray will not be possible. The issue occurs sporadically; however, it may occur during an on-going procedure.
Firm Action
Urgent Medical Device Correction notification letters dated 6/12/19 were sent to customers.
In the event the issue occurs, the system cannot be operated normally as X-ray will not possible. Planned procedures may have to be terminated and performed on an alternative X-ray system. Siemens advises that an alternative system is available to continue treatment in case of occurrence.
Siemens is implementing a corrective action to all potentially affected users of the Artis Zee and AXIOM Artis systems with generator A100 via Update Instruction AX038/18/S. The Siemens Service organization will modify the affected supply circuit and revise the wiring. This modification will bring the voltage back to the center of the tolerance range.
Our service organization will contact you shortly to arrange a date to perform this corrective action. Please feel free to contact our service organization for an earlier appointment at 1-800-888-7436. Following the correction, the cause is eliminated and any recurrence of this potential fault is prevented.
Root Cause
Component design/selection
Distribution Pattern
US Nationwide Distribution
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.
Advertisement