INTUA CRT-P, Model Numbers: a) V272 b) V273 c) W273
FDA Recall Z-2019-2021 · Posted 2021-07-13
Advertisement
Recalling FirmBoston Scientific Corporation
Recall StatusOpen, Classified
Product CodeNKE
Date Initiated2021-06-03
Product Quantity1707 units
Reason for Recall
The dual chamber INGENIO family pacemakers or cardiac resynchronization therapy pacemakers (CRT-Ps) may initiate Safety Mode later in device life (i.e., prior to reaching the Explant battery indicator) when the device's battery exhibits high internal impedance. This latent battery condition puts a device at risk for system resets to occur due to temporary high-power consumption related to telemetry attempts and subsequent reversion to Safety Mode to maintain back-up pacing. Although therapy is still provided when a device is in Safety Mode, replacement is required.
- Only dual chamber INGENIO family pacemakers and CRT-Ps built with the Extended Life (EL) battery are included within this advisory population
Firm Action
The firm issued an advisory communication on June 3, 2021 by letter to the implanting/following physicians and clinics regarding the potential device behavior. The notice included a letter to the patients to be disseminated to the patients implanted with the device at the discretion of the physician. BSC also made the patient letter available on their advisory page, found on our product performance website @ https://www.bostonscientific.com/en-US/pprc/product-advisories.html
Root Cause
Device Design
Distribution Pattern
Worldwide distribution.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.