Intuitive Surgical Inc. Large Clip Applier instruments used in conjunction with da Vinci S and da Vinci SI Surgical Systems Intuitive Surgical Inc., 'Sunnyvale, CA 94086
FDA Recall Z-2020-2014 · Posted 2014-07-08
Advertisement
Recalling FirmIntuitive Surgical, Inc.
Recall StatusTerminated
Product CodeNAY
Date Initiated2014-06-13
Date Terminated2014-12-01
Product Quantity5290 for part 420230-04; 1964 for part 420240-06
Reason for Recall
Manufacturing variation in the grip assembly of the Large Clip applier which can lead to the instrument jamming during insertion when used with the 8 mm cannula or with certain reducer accessories..
Firm Action
Urgent Medical Device Correction letters were sent on June 18, 2014 to all affected sites via trackable mail.
Root Cause
Process control
Distribution Pattern
Worldwide Distribution.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.