Terminated

Intuitive Surgical Inc. Large Clip Applier instruments used in conjunction with da Vinci S and da Vinci SI Surgical Systems Intuitive Surgical Inc., 'Sunnyvale, CA 94086

FDA Recall Z-2020-2014 · Posted 2014-07-08

Recall StatusTerminated
Product CodeNAY
Date Initiated2014-06-13
Date Terminated2014-12-01
Product Quantity5290 for part 420230-04; 1964 for part 420240-06

Reason for Recall

Manufacturing variation in the grip assembly of the Large Clip applier which can lead to the instrument jamming during insertion when used with the 8 mm cannula or with certain reducer accessories..

Firm Action

Urgent Medical Device Correction letters were sent on June 18, 2014 to all affected sites via trackable mail.

Root Cause

Process control

Distribution Pattern

Worldwide Distribution.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑