AEQUALIS FLEX REVIVE Assembly Screw 0mm Short, Model Number ARS655118

FDA Recall Z-2020-2023 · Posted 2023-06-28

Recalling FirmTornier, Inc
Recall StatusOpen, Classified
Product CodeKWS
Date Initiated2023-06-12
Product Quantity55 units

Reason for Recall

A comingle/swap resulted in the labeling and laser marking on the product to state it is a Aequalis Flex Revive Assembly Screw 0mm Standard (ARS655101) but it was a Aequalis Flex Revive Assembly Screw 0mm Short (ARS655118) & vice versa.

Firm Action

Tornier issued an URGENT: MEDICAL DEVICE RECALL notice to its consignees beginning on 06/12/2023 by email and letter. The notice explained the problem, the risk associated with its use, and requested that the affected devices be isolated/quarantined pending return. Distributors were directed to notify their customers.

Root Cause

Under Investigation by firm

Distribution Pattern

Worldwide distribution - US Nationwide and the countries of Australia, Canada, Netherlands, Spain, United Kingdom.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.