Terminated

STOPCOCK EXT. SET W/ 3-WAY STOPCOCK WITH INJ. SITE AND SPINLOCK CONNECTOR, 22 inches and 37 inches

FDA Recall Z-2022-2014 · Posted 2014-07-09

Recall StatusTerminated
Product CodeFPA
Date Initiated2014-05-29
Date Terminated2014-12-10
Product Quantity10,835

Reason for Recall

Complaints were received that reported the stopcock of some Stopcock Extension Sets are assembled in the reverse position, resulting in the stopcock being closed when set in the open position and vise versa.

Firm Action

An urgent medical device correction notification, dated June 3, 2014, describing the product, problem, and actions to be taken, was sent to end users and distributors.

Root Cause

Process control

Distribution Pattern

USA (nationwide) including the states of PA, IL, NC, FL, IL, CA, CO, NE, LA, NC, LA, VA, TN, NE, TX, and KS.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →