STOPCOCK EXT. SET W/ 3-WAY STOPCOCK WITH INJ. SITE AND SPINLOCK CONNECTOR, 22 inches and 37 inches
FDA Recall Z-2022-2014 · Posted 2014-07-09
Recalling FirmB. Braun Medical, Inc.
Recall StatusTerminated
Product CodeFPA
Date Initiated2014-05-29
Date Terminated2014-12-10
Product Quantity10,835
Reason for Recall
Complaints were received that reported the stopcock of some Stopcock Extension Sets are assembled in the reverse position, resulting in the stopcock being closed when set in the open position and vise versa.
Firm Action
An urgent medical device correction notification, dated June 3, 2014, describing the product, problem, and actions to be taken, was sent to end users and distributors.
Root Cause
Process control
Distribution Pattern
USA (nationwide) including the states of PA, IL, NC, FL, IL, CA, CO, NE, LA, NC, LA, VA, TN, NE, TX, and KS.