ev3 The Endovascular Company, Protege GPS 8mm x 40mm Self Expanding Peripheral Stent System , Rx Only, and SERP65-08-40-120. In the US, the intended as a palliative treatment of malignant neoplasms in the biliary tree. Outside the US, the intended for use in occlusions, lesions at high risk for abrupt closure or threatened closure following percutaneous transluminal angioplasty (PTA) or lesions believed to be at high risk for restenosis following PTA in the common and external iliac, subclavian or renal arteries.
FDA Recall Z-2025-2009 · Posted 2009-09-08
Advertisement
Recalling FirmEv3, Inc.
Recall StatusTerminated
Product CodeFGE
Date Initiated2009-06-10
Date Terminated2011-12-11
Product Quantity2
Reason for Recall
ev3 Inc. is conducting a voluntary recall of specific lots of the Protege GPS Self-Expanding Stent System because labeling did not match the actual stent size contained in the package. The affected product was labeled as a 6mm x 80mm Protege GPS stent but actually contained an 8mm x 40mm stent. This situation can lead to vessel damage or inadequate coverage of the target lesion.
Firm Action
Consignees were emailed a letter with a subject heading of "Medical Device Recall". The e-mail was dated June 10, 2009. The letter informed the consignee of the problem and product involved. They requested that any affected devices remaining in inventory be identified and returned and to fill out and return the "Device Recall Field Action Reconciliation Form".
Root Cause
Labeling mix-ups
Distribution Pattern
International Distribution Only -- Brazil, Uruguay, & Malaysia.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.