OEC 9800, OEC FluoroTrak 9800 Plus, OEC 9800 Plus, OEC 9800 MD Motorized C-arm System designed to provide fluoroscopic and spot-film imaging of the patient during diagnostic, surgical and interventional procedures. The OEC 9800 is designed to provide fluoroscopic and spot-film imaging of the patient during diagnostic, surgical and interventional procedures. The system includes features specifically designed for use in diagnostic and interventional cardiac imaging procedures, and is also intended for cholangiography, endoscopic, urologic, orthopedic, neurologic, vascular, critical care and emergency room procedures. It may be used for other imaging applications at the physician's discretion.
FDA Recall Z-2026-2012 · Posted 2012-07-18
Recalling FirmGE OEC Medical Systems, Inc
Recall StatusTerminated
Product CodeJAA
Date Initiated2007-11-19
Date Terminated2013-01-28
Product Quantity11,100 units
Reason for Recall
Please be aware that this is not a new recall. The firm has taken action; but, due to administrative issues this recall is now being classified by the Agency. GE OEC recalled imaging devices OEC 9900, OEC 9800, and OEC 8800 as a result of an FDA inspection identifying that the vertical lift column power supply in the mainframe C-arm is defective and subject to early life failure.
Firm Action
Please be aware that this is not a new recall. The firm has taken action; but, due to administrative issues this recall is now being classified by the Agency. GE Healthcare sent an"URGENT RECALL NOTICE" dated February 22, 2008 to all affected customers. The letter identifies the product, problem, and actions to be taken by the customers. Contact the firm at 800-874-7378 for questions concerning this recall.
Root Cause
Device Design
Distribution Pattern
Worldwide Distribution.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.