Terminated

Medtronic StealthStation S7 System

FDA Recall Z-2029-2010 · Posted 2010-07-17

Recall StatusTerminated
Product CodeHAW
Date Initiated2010-06-14
Date Terminated2011-02-18
Product Quantity370 units

Reason for Recall

Excessive wear at the power cord retractor or incorrect assembly of the power plug may cause sparking, intermittent loss of power, or electrical shock.

Firm Action

Customers were notified by letter on Jun 14, 2010, advising them of the problem. Instructions were provided to continue using existing retractable power cords until they could be replaced with non-retractable cords by Medtronic, beginning on Aug 17. Further information is available from the company by calling 800-595-9709.

Root Cause

Device Design

Distribution Pattern

Nationwide, including 3 VA facilities. No military distribution. Foreign accounts in Australia, Austria, Belgium, Canada, Chile, Czech Republic, Denmark, Egypt, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Indonesia, Ireland, Italy, Latvia, Libya, Lithuania, Luxembourg, Malaysia, Namibia, Netherlands, New Zealand, Poland, Puerto Rico, Russia, Saudi Arabia, Singapore, Slovakia, South Africa, South Korea, Spain, Sweden, Switzerland, Thailand, Turkey, United Kingdom, United Arab Emirates.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.