Sarns Disposable Centrifugal Pump with X-Coating, 164275X, is a sterile, single use device. For use in cardiopulmonary bypass procedures.
FDA Recall Z-2029-2014 · Posted 2014-07-11
Advertisement
Recalling FirmTerumo Cardiovascular Systems Corporation
Recall StatusTerminated
Product CodeKFM
Date Initiated2014-04-14
Date Terminated2016-09-07
Reason for Recall
During set-up and priming of the bypass circuit, leaks were detected on the centrifugal pumphead.
Firm Action
Urgent Medical Device Recall letters, dated April 11, 2014, were sent to customers. The letters identified the affected product, as well as the reason for the correction, the potential hazard, the correction, the affected population, and instructions for customers. Customers are to review the notice and ensure that all users have received a copy. Additionally, customers are to confirm receipt of the communication by returning the attached Customer Response Form as instructed; and, return all effected centrifugal pumps. Terumo CVS will issue a Returned Goods Authorization upon receipt of the Response Form. Questions should be directed to Terumo CVS Customer Service at 800-521-2818.
Root Cause
Device Design
Distribution Pattern
Nationwide Distribution