Destination Renal Guiding Sheath Intended for the introduction of interventional and diagnostic devices into the human vasculature.

FDA Recall Z-2033-2017 · Posted 2017-05-04

Recall StatusTerminated
Product CodeDYB
Date Initiated2015-11-06
Date Terminated2017-05-12
Product Quantity5,252 eaches

Reason for Recall

Based on internal investigation and testing that revealed the distal end of certain lots of Terumo Medical Destination products may not contain the labeled 5 cm of coating. While there have been no complaints or patient related incidents reported, the lack of coating could render it difficult to navigate the device to the target vasculature.

Firm Action

An Urgent Product Bulletin Recall of Terumo Medical Product letter, dated November 6, 2015, was sent to customers. The letter identified the affected product and the reason for the voluntary recall. The letter asked customers to immediately discontinue use and quarantine any of the affected product listed. Customers are to complete the enclosed Contact & Inventory Confirmation Form and e-mail [email protected] or fax to 732-412-4121, Attn: Destination Recall. For questions or further information, customers can contact Customer Care at 800-888-3786 or their local sales representative.

Root Cause

Other

Distribution Pattern

Nationwide Distribution.