BI-700-00027 0-arm 1000 Imaging System Mobile X-ray System Generating 2D and 3D images of human anatomy for surgical applications.
FDA Recall Z-2034-2008 · Posted 2008-09-26
Advertisement
Recalling FirmMedtronic Navigation, Inc.
Recall StatusTerminated
Product CodeJAA
Date Initiated2008-06-30
Date Terminated2008-12-04
Product Quantity10 units
Reason for Recall
Failure to of affected units to comply with EER/AKR limits due to misinterpretation of the measurement requirements specified in 21 CFR 1020.32 (d) (3) (iii)
Firm Action
Medtronic Navigation notified service engineers to retest units for compliance.
Root Cause
Radiation Control for Health and Safety Act
Distribution Pattern
CA, DC, LA, FL. IN, OH, PA, VA