Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Terminated

NC Trek RX Coronary Dilatation Catheter, Part No. 1012448-06, 1012448-08, 1012448-12, 1012448-15, 1012448-20, 1012449-06, 1012449-08, 1012449-12, 1012449-15, 1012449-20, 1012449-25, 1012450-06, 1012450-08, 1012450-12, 1012450-15, 1012450-20, 1012451-06, 1012451-08, 1012451-12, 1012451-15, 1012451-20, 1012451-25, 1012452-08, 1012452-12, 1012452-15, 1012452-20, 1012453-08, 1012453-12, 1012453-15, 1012453-20

FDA Recall Z-2037-2017 · Posted 2017-05-12

Recalling FirmAbbott Vascular
Recall StatusTerminated
Product CodeLOX
Date Initiated2017-03-22
Date Terminated2018-02-07
Product Quantity449,661 units total (132,040 units in US)

Reason for Recall

Products from the identified lots may exhibit difficulty in removing the protective balloon sheath which can result in issues with inflating or deflating the balloon.

Firm Action

An Urgent Field Safety Notice dated 3/22/17 was sent to customers to inform them that Abbott Vascular is recalling the NC Trek RX Coronary Dilation Catheter, NC Traveler RX Coronary Dilatation Catheter, and NC Tenku RX PCTA Balloon Catheter. Products from the identified lots may exhibit difficulty in removing the protective balloon sheath which can result in issues with inflating or deflating the balloon. The letter informs the customers of how does the issue occur and what action is Abbott Vascular asking the customers to take. Customers with any questions are instructed to contact local Abbott Vascular Representative or Customer Service Department at (800) 227-9902. On 5/15/17, Abbott Vascular posted a press release on their website to inform customers that Abbott has initiated a voluntary recall of specific lots of three catheters: NC Trek RX Coronary Dilatation Catheter, NC Traveler Coronary Dilatation Catheter, and NC Tenku RX PTCA Balloon Catheter. Products from the identified lots may exhibit difficulty in removing the protective balloon sheath, which could cause problems with inflating or deflating the balloon. Potential risks associated with balloon inflation and deflation difficulties include air embolism, additional intervention, thrombosis, and myocardial infarction.

Root Cause

Device Design

Distribution Pattern

US and Worldwide

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑