NC Traveler RX Coronary Dilatation Catheter, Part No. 1013152-12, 1013152-15, 1013152-20, 1013153-08, 1013153-12, 1013153-15, 1013153-20, 1013154-08, 1013154-12, 1013154-15, 1013154-20, 1013155-08, 1013155-12 1013155-15, 1013155-20, 1013156-12, 1013156-15, 1013157-08, 1013157-12, 1013157-15
FDA Recall Z-2038-2017 · Posted 2017-05-12
Advertisement
Recalling FirmAbbott Vascular
Recall StatusTerminated
Product CodeLOX
Date Initiated2017-03-22
Date Terminated2018-02-07
Product Quantity449,661 units total (132,040 units in US)
Reason for Recall
Products from the identified lots may exhibit difficulty in removing the protective balloon sheath which can result in issues with inflating or deflating the balloon.
Firm Action
An Urgent Field Safety Notice dated 3/22/17 was sent to customers to inform them that Abbott Vascular is recalling the NC Trek RX Coronary Dilation Catheter, NC Traveler RX Coronary Dilatation Catheter, and NC Tenku RX PCTA Balloon Catheter. Products from the identified lots may exhibit difficulty in removing the protective balloon sheath which can result in issues with inflating or deflating the balloon. The letter informs the customers of how does the issue occur and what action is Abbott Vascular asking the customers to take. Customers with any questions are instructed to contact local Abbott Vascular Representative or Customer Service Department at (800) 227-9902.
On 5/15/17, Abbott Vascular posted a press release on their website to inform customers that Abbott has initiated a voluntary recall of specific lots of
three catheters: NC Trek RX Coronary Dilatation Catheter, NC Traveler Coronary Dilatation Catheter, and NC Tenku RX PTCA Balloon Catheter. Products from the identified lots may exhibit difficulty in removing the protective balloon sheath, which could cause problems with inflating or deflating the balloon. Potential risks associated with balloon inflation and deflation difficulties include air embolism, additional intervention, thrombosis, and myocardial infarction.
Root Cause
Device Design
Distribution Pattern
US and Worldwide
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.
Advertisement